Microprocessor equipped NDIR Measuring Sensor Boards
CO2 flow sensor using single channel technology
IR sensors using one channel technology. Optional in IP 54 ABS case. Diffusion- or Pump version. Available for many measuring ranges from 2000 ppm to 100 % for a variety of OEM applications.
Properties
Methane flow sensor using single channel technology
  • Using microprocessor technology, all on one board
  • Short response time (T90 < 45 seconds)
  • Zeroing of the sensor: manually over zero button or externally over contacts
  • Automatic temperature compensation
Working modes
Pictures
  • :"Smart Zero" - automatic correction of sensor zero using values for the past 1 to 30 days
  • Continuous measurement - permanent measuring mode (can be zeroed externally)
Product info
Brochure (50 kB)
Manual (355 kB))
Control software (25 MB)
Standard configuration
  • Linear voltage output - active range (begin and end) may be user-configured within the nominal range 0....10 V.
  • Linear current output - active range (begin and end) may be user-configured within the nominal range 4....20 mA
  • Interface RS232C for changing the sensor settings and for data-logging
  • Digital PWM output (0/5 V) - programmable output
  • Sensor zeroing - manually or over contacts on the board
  • Power supply 13...30 VDC or 9...24 VAC, max 3 W
Optional equipment
  • Membrane pump for sample collection or ventilation
  • USB connection over optional adapter
Preparation and display of measuring values
  • Measured values in digital form over RS232C
  • Measured values over analogue outputs: 0/4-20 mA and 0-10 V
  • PC programme for changing the sensor settings and for data-logging>
Operating data

Parameter

madIR-D01
Measured gases
CO, CO2, CH4
Measuring method
infra-red absorption, 1 channel with light modulation
Sampling flow
Operating temperature
0 ÷ +50°C
Storage temperature
-20 ÷ +70°C
Humidity
5 ÷ 90 %, non-condensing
Voltage supply
13...30 VDC or 9...24 VAC
Power
max. 3 W
Analogue outputs

Begin and end can be user-configured within the range
Current: 4 - 20 mA, linear
Voltage
: 0 - 10 V, linear

Interface RS 232C
Averaging time
2 ÷ 60 s user-configured
Calibration
10 point calibration stored in EEPROM
Recalibration
not necessary. Falls erwünscht 2-Punkte Kalibrierung möglich
Size of Printplatte

105 x 40 x 30 mm

Weight 120 g

Measuring Technical Data
Gas Range

Resolution

Detection limit Accuracy Resp. time (t90)
CO - carbon monoxide-volumetric concentration

CO2 - carbon dioxide-volumetric concentration

CH4 - methane-volumetric concentration
0...100 % 0.10 % 0.10 % ± 0.5 % abs., or 3 % rel. 45 s
0....50 % 0.10 % 0.10 % ± 0.3 % abs., or 3 % rel.
0...25 % 0.01 % (100 ppm) 0.01 % (100 ppm) ± 0.15 % abs., or 3 % rel.
0...10 % 0.01 % (100 ppm) 0.01 % (100 ppm) ± 0.05 % abs., or 3 % rel.
0...5 % 0.01 % (100ppm) 0.01 % (100ppm) ± 0.03 % abs., or 3 % rel.
0...2.5 % 0.001 % (10 ppm) 0.001 % (10 ppm) ± 0.015 % abs., or 3 % rel.
a