madIR-D01
OEM sensor
Microprocessor equipped NDIR Measuring Sensor Boards
|
|
 |
|
| IR sensors using one channel technology. Optional in IP 54 ABS case. Diffusion- or Pump version. Available for many measuring ranges from 2000 ppm to 100 % for a variety of OEM applications. |
|
|
|
|
 |
- Using microprocessor technology, all on one board
- Short response time (T90 < 45 seconds)
- Zeroing of the sensor: manually over zero button or externally over contacts
- Automatic temperature compensation
|
|
|
|
|
|
|
|
|
|
|
- :"Smart Zero" - automatic correction of sensor zero using values for the past 1 to 30 days
- Continuous measurement - permanent measuring mode (can be zeroed externally)
|
|
|
|
|
|
|
-
- Linear voltage output - active range (begin and end) may be user-configured within the nominal range 0....10 V.
- Linear current output - active range (begin and end) may be user-configured within the nominal range 4....20 mA
- Interface RS232C for changing the sensor settings and for data-logging
- Digital PWM output (0/5 V) - programmable output
- Sensor zeroing - manually or over contacts on the board
- Power supply 13...30 VDC or 9...24 VAC, max 3 W
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Preparation and display of measuring values
|
|
|
- Measured values in digital form over RS232C
- Measured values over analogue outputs: 0/4-20 mA and 0-10 V
- PC programme for changing the sensor settings and for data-logging>
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
madIR-D01 |
|
Measured gases
|
CO, CO2, CH4 |
| Measuring method |
infra-red absorption, 1 channel with light modulation
|
| Sampling |
flow |
|
Operating temperature
|
0 ÷ +50°C |
|
Storage temperature
|
-20 ÷ +70°C |
|
Humidity
|
5 ÷ 90 %, non-condensing |
|
Voltage supply
|
13...30 VDC or 9...24 VAC |
|
Power
|
max. 3 W |
|
Analogue outputs
|
Begin and end can be user-configured within the range
Current: 4 - 20 mA, linear
Voltage: 0 - 10 V, linear
|
| Interface |
RS 232C |
|
Averaging time
|
2 ÷ 60 s user-configured |
|
Calibration
|
10 point calibration stored in EEPROM |
|
Recalibration
|
not necessary. Falls erwünscht 2-Punkte Kalibrierung möglich |
| Size of Printplatte |
105 x 40 x 30 mm
|
| Weight |
120 g |
|
Measuring Technical Data
|
| Gas |
Range |
Resolution
|
Detection limit |
Accuracy |
Resp. time (t90) |
CO - carbon monoxide-volumetric concentration
CO2 - carbon dioxide-volumetric concentration
CH4 - methane-volumetric concentration |
0...100 % |
0.10 % |
0.10 % |
± 0.5 % abs., or 3 % rel. |
45 s |
| 0....50 % |
0.10 % |
0.10 % |
± 0.3 % abs., or 3 % rel. |
| 0...25 % |
0.01 % (100 ppm) |
0.01 % (100 ppm) |
± 0.15 % abs., or 3 % rel. |
| 0...10 % |
0.01 % (100 ppm) |
0.01 % (100 ppm) |
± 0.05 % abs., or 3 % rel. |
| 0...5 % |
0.01 % (100ppm) |
0.01 % (100ppm) |
± 0.03 % abs., or 3 % rel. |
| 0...2.5 % |
0.001 % (10 ppm) |
0.001 % (10 ppm) |
± 0.015 % abs., or 3 % rel. |