Products >> Gas Analysers >> Portable Analysers >> GA-40 Tplus
GA-40Tplus portable flue gas analyser with integrated sample conditioner, here shown without heated hose
GA-40Tplus

Portable flue gas analyser using electrochemical and IR gas sensors.

The flue gas analyser GA-40Tplus is designed especially for service technicians who are permanently employed in measuring emissions and carrying out adjustments as well as making long-term measurements and inspections on various burner and heating equipment. For this purpose the analyser is fitted with a high-powered dryer and heated hose. The analyser can be equipped with 9 sensors (7 electrochemical and 2 NDIR sensors).

All parts of the analyser are designed for a long and trouble-free operation. 6 additional analogue inputs for voltage/current signals and temperature together with 2 optional analogue outputs give the instrument an unheard of flexibility. The analyser can also be fitted with two IR sensors for direct measurement of CO2 and CH4. Constructed to be compliant with EN50379.

Pictures
Product info
Brochure (68 kB)
Manual (605 kB)
PC software (17,6 MB)
Standard configuration
  • Measurement of gas concentrations
    • Gas concentrations measured with NDIR sensors: CO2, CH4
    • Gas concentrations measured with electrochemical sensors: O2
    • Toxic gas concentrations measured with electrochemical sensors: CO, NO, NO2, SO2, H2S, H2. Please ask for quote for other gases.
  • Measurement of other parameters
    • Measurement of gas and ambient temperatures
    • Pressure, draft and differential pressure measurements with 1 Pa resolution
    • Soot test according to Bacharach with a pump flow of 1.62 l/min
    • 6 analogue inputs (2 current/voltage, 2 thermocouple, 2 thermistor)
  • Calculation
    • CO2 concentration (if no NDIR sensor for CO2 is fitted)
    • Calculation of absolute and relative mass concentrations & relative emissions
    • Calculation of all relevant combustion parameters
  • Processing and presentation of measured data
    • All results shown on display
    • Averaging of all measured values. Averaging time: 10 sec - 60 min
    • Graphical presentation of all measured values as diagram
    • Memory capacity for 1024 sets of data, organised into 10 data banks
    • Up to 9 separate items in a data set
    • User definition of the 9 items
    • Memory capacity for 30 reports
    • Data logger function for the analogue inputs
    • All measured values, stored values or displays can be printed on the internal printer.
    • Powerful PC programme for analyser settings and data communication
  • Software-Eigenschaften
    • International compatibility (language, date format etc.)
    • Password protection for settings
    • Automatic zeroing when the analyser is switched on
    • Calibration of O2/CO2 during use
    • All parameters programmable
    • List of 22 common fuels
    • 10 further freely programmable fuels
    • Permanent automatic check of the instruemnt with acoustic warning and full information in the "control list"
    • Compensation of cross sensitivity and temperature drift of gas sensors
  • Hardware capabilities
    • Peltier Gas dryer
    • CO measurement - separate from the other gas channels. If the freely programmable maximum is exceeded, then the sensor is purged with air, without interrupting the other measurements
    • Electronic regulation of the pump flow rate
    • Integrated clock/calendar
    • Internal 57 mm dot-matrix printer
    • Power from rechargeable battery or mains
    • Big (70 x 60 mm) LCD display with backlighting (Graphics or 11 x 21 characters)
    • Gas probe with thermocouple and condensate trap
    • Power for heated probe holder during mains operation
    • RS-232C interface and multifunctional PC programme
Optional accessories
    • External ambient temperature sensor
    • Remote control
    • Flow velocity
Technical data
    • Size of case W x H x D: 485 x 205 x 295 mm
    • Weight w/o probe: 10 kg
    • Display size: Graphical - LCD, with backlighting and variable contrast, 128 x 112 pixels, 75 x 64 mm
    • Printer: high-speed dot matrix printer with graphics capability for 57 mm normal paper
    • Data memory: Memory for 30 reports and 10 banks containg a total of 1024 sets of data 
    • CO - measurement channel: Separate from the other gas channels. If the freely programmable maximum is exceeded then the sensor is purged with air, without interrupting the other measurements
    • Interface: RS232C
    • Power supply: 110/230 VAC 50 ÷ 60 Hz
    • Charging time: Lead-acid battery 12V / 2.2 Ah, charging time 10 h, working time approx. 6 h (w/o dryer)
    • Gas pump: Membrane pump, electronically regulated at 90l/h
    • Probe length: 300 mm (other options available)
    • Length of heated line: 3m
    • Peltier dryer
    • Gas filter: In line filter 20 µm
    • Operating temperature: 10 °C ÷ 50°C
    • Storage temperature: -20 °C ÷ +55 °C
Parameter
Method
Indication range
Display resolution
Accuracy
Detection limit
Resp. time (t90)
Gases measured in standard configuration
O2 - oxygen, volumetric concentration
electrochemical gas sensor
0..25 %
0.01 %
±0.2% or
2% rel.
0.20 %
45 s
CO2 - carbon dioxide, volumetric concentration
calculated from volumetric concentration of O2 and fuel
0..25 %
0.01%
±0.2% or
2% rel.
0.20 %
45 s
CO - carbon monoxide, volumetric concentration
electrochemical gas sensor
0..20000 ppm
1 ppm
±5 ppm or
5 % rel.
5 ppm
45 s
COmg - carbon monoxide, mass concentration
calculated from volumetric concentration of CO
0...
1 mg/Nm3
±10 mg/Nm3 or 5 % rel.
10 mg/Nm3
45 s
COrel - carbon monoxide, mass concentration relative to O2
calculated from volumetric concentration of CO and O2
0...
1 mg/Nm3
±10 mg/Nm3 or 5 % rel.
10 mg/Nm3
45 s
Gases measured with optional IR sensors
CO2 - carbon dioxide volumetric concentration
NDIR sensor
0..25 %
0...100 %
0.01 %
+/-3 % rel., or 0.5 % abs.
0.20 %
45 s
CH4 - Methane, volumetric concentration
NDIR sensor
0...5 %
0...100 %
0.01 %
0.1 %
+/-3 % rel., or 0.5 % abs.
0.20 %
45 s
Gases measured with optional electrochemical sensors
NO / NOx - volumetric concentration of nitrogen oxides.
electrochemical gas sensor
0..1000 ppm
1 ppm
±5 ppm or 5 % rel.
5 ppm
45 s
NOmg/NOxmg - mass concentration of nitrogen oxides
calculated from volumetric concentration of NO
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15 mg/Nm3
45 s
NOrel / NOxrel - mass concentration of nitrogen oxides relative to O2
calculated from volumetric concentration of NO and O2
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15 mg/Nm3
45 s
NO2 - volumetric concentration of nitrogen dioxide
electrochemical gas sensor
0..1000 ppm
1 ppm
±5 ppm or 5 % rel.
5 ppm
45 s
NO2mg - mass concentration of nitrogen dioxide
calculated from volumetric concentration of NO2
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15 mg/Nm3
45 s
NO2rel - mass concentration of nitrogen dioxide relative to O2
calculated from volumetric concentration of NO2 and O2
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15 mg/Nm3
45 s
SO2 - volumetric concentration of sulphur dioxide
electrochemical gas sensor
0..5000 ppm
1 ppm
±5 ppm or
5 % rel.
5 ppm
45 s
SO2mg - mass concentration of sulphur dioxide
calculated from volumetric concentration of SO2
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15mg/Nm3
45 s
SO2rel - mass concentration of sulphur dioxide relative to O2
calculated from volumetric concentration of SO2 and O2
0...
1 mg/Nm3
±15 mg/Nm3 or 5 % rel.
15 mg/Nm3
45 s
Other measured values
Tgas - flue gas temperature
Thermocouple
-10..1000°C
1°C
± 2 °C or 1.5 % rel.
1 °C
30 s
Tamb - ambient temperature
Thermistor
-10..100°C
1°C
± 1 °C
1 °C
30 s
U1,U2 - 2 external inputs
ADC
-20...+20 V
0.01 V
±0.02 V
0.01 V
10 s
I1,I2 - 2 external inputs
ADC
-20...+20 mA
0.01 mA
±0.02 mA
0.01 mA
10 s
T1,T3 - 2 external inputs
Thermocouple
0...1600°C
1°C
±2 °C or 1. 5 % rel.
1 °C
10 s
T2,T4 - 2 external inputs
Thermistor
-20...100°C
1°C
±2 °C or 1. 5 % rel.
1 °C
10 s
Pressure
DMS bridge
-25 hPa ... +25 hPa
0.1Pa
±2 Pa or 5 % rel.
1 Pa
10 s
Flow velocity (Option)
Pitot tube
1...50 m/s
0.1 m/s
0.3 m/s or 5 % rel.
0.1 m/s
10 s
Soot/smoke test
Bacharach method
0..9
0.5
0.5
0.5
TI (CO/CO2-Toxic Index)
calculated
0...0.01
0.0001
5 % rel.
0
10 s
Lambda - excess air number
calculated
1..10
0.01
2 % rel.
0
10 s
qA - combustion losses
calculated
0..100 %
0.1 %
2 % rel.
0 %
10 s
Eta - efficiency
calculated
0..120 %
0.1 %
2 % rel.
0 %
10 s