GA-60
Portable flue gas analyser using electrochemical and IR gas sensors. |
|
|
The GA-60 is a high-quality instrument produced with use of the latest technical discoveries. In addition to its exceptional measurement abilities it is also handy and easy to use.
The instrument is fitted with 5 electrochemical sensors as standard. A further 2 electrochemical sensors and 3 infrared sensors can be added. The modular construction allows the instrument to be configured for nearly every customer requirement. The instrument also has various analog inputs and outputs to enable easy data transfer and documentation.Constructed to be compliant with EN50379.
|
|
|
|
|
|
|
|
|
|
|
- Measurement of gas concentrations
- Concentrations measured with electrochemical sensors: O2, CO, NO, NO2, SO2.
- Optional two electrochemical sensors: H2S, H2
- Gas concentrations measured with NDIR sensors: N2O, CO2, CH4
- Measurement of other parameters
- Measurement of gas and ambient temperatures
- Pressure, draft and differential pressure measurements with 1 Pa resolution
- Ambient pressure measurement with resolution of 0.1 hPa option
- 8 Temperature channels (4 thermocouples and 4 thermistors) option.
- Soot test according to Bacharach with a pump flow of 1.63 l/min
- Calculation
- CO2 concentration if no CO2 sensor is fitted.
- Calculation of all relevant combustion parameters.
- Calculation of absolute and relative mass concentrations for the measured gases
- Calculation of absolute and relative volumetric concentrations for the measured gases
- Processing and presentation of measured data
- All measured and calculated values can be displayed as averaged values as well. Averaging time can be chosen from the series: 2, 10, 20, 30, 60, 120, and 180 seconds.
- Single and triple long-term measurements (XL measurements). Period for long-term measurements chosen from the series: 10 s, 20 s, 30 s, 1 min., 5 min., 10 min., 15 min., 20 min., 30 min.
- Single or continuous storage of results. One set of data will contain all measured and calculated values.
- Memory for 1M sets of data (MMC 256MB)
- Complete software package for the PC to process readings and communicate on-line.
- Software features
- Automatic zero calibration on switch-on
- All parameters can be freely programmed
- Complete list of 10 standard fuels
- Freely programmable fuels
- Continuous automatic monitoring of instrument function with acoustic warning and detailed information under “Control List”
- Cross-sensitivity and temperature drift of gas sensors is fully compensated
- Hardware features
- Data logger with full 256 MB memory to facilitate continuous storing of results. The results can be read from the memory into a PC.
- Flash memory for all instrument settings
- Large LCD display (320 x 240) with backlighting
- Internal dot matrix printer 57 mm wide
- PC interface RS232C for communication with the analyser and data transfer
- Power for heated probe holder during mains operation
- Power from rechargeable battery or mains
- Twin gas filters (20 µm, 5 µm) and rotameter integrated in gas system
- Integral clock/calendar with separate buffer battery
|
|
|
|
|
- 8 Temperature channels (4 thermocouples and 4 thermistors)
- Output signal module. Freely programmable analogue outputs (8 channels voltage 0…11 V/12 bit resolution and 8 channels current 0…25 mA/12 bit resolution). Can be set to any parameter.
|
|
|
|
|
- Size of case: BxTxH: 500 x 410 x 180 mm
- Weight w/o probe: 10 kg
- Display size: Graphical - LCD, with backlighting , 320 x 240 Pixel
- Printer: high-speed dot matrix printer with graphics capability for 57 mm normal paper
- Data memory: 256 MB memory to facilitate continuous storing of results
- CO - measurement channel: Separate from other gas channels. If the programmed maximum is exceeded, then the CO sensor will be automatically purged without inhibiting the other measurement channels.
- Interface: RS232C
- Power supply: 110/230 VAC 50 ÷ 60 Hz
- Charging time: Lead-acid battery 12V / 2.2Ah, charging time 10 h, working time approx. 6h
- Gas pump: Membrane pump
- Probe length: 300 mm (other options available), heated for soot test
- Length of gas linee: 3m
- Gas filter: in line filter 5µm and 20 µm
- Operating temperature: 10 °C ÷ 50°C
- Storage temperature: -20 °C ÷ +55 °C
- Ambient humidity 5 90 %, non-condensing
|
|
|
|
|
|
|
|
| Parameter |
Method
|
Indication range
|
Display resolution
|
Accuracy
|
Detection limit
|
Resp. time (t90)
|
|
Gases measured in standard configuration |
| O2 - oxygen, volumetric concentration |
electrochemical gas sensor
|
0..25 %
|
0.01 %
|
±0.2% or
2% rel. |
0.20 % |
45 s
|
| CO2 - carbon dioxide, volumetric concentration |
calculated from volumetric concentration of O2 and fuel
|
0..25 %
|
0.01%
|
±0.2% or
2% rel. |
0.20 % |
45 s
|
| CO - carbon monoxide, volumetric concentration (optional) |
electrochemical gas sensor
|
0...10 %
|
10 ppm |
±50 ppm or
5 % rel.
|
10 ppm
|
45 s
|
| CO - carbon monoxide, volumetric concentration |
electrochemical gas sensor
|
0..20000 ppm
|
1 ppm |
±5 ppm or
5 % rel.
|
5 ppm
|
45 s
|
| COmg - carbon monoxide, mass concentration |
calculated from volumetric concentration of CO
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
1 mg/Nm3
|
45 s
|
| COrel - carbon monoxide, mass concentration relative to O2 |
calculated from volumetric concentration of CO and O2
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
1 mg/Nm3
|
45 s
|
| NO / NOx - volumetric concentration of nitrogen oxides. |
electrochemical gas sensor
|
0..5000 ppm
|
1 ppm
|
±5 ppm or 5 % rel.
|
5 ppm
|
45 s
|
| NOmg/NOxmg - mass concentration of nitrogen oxides |
calculated from volumetric concentration of NO
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
1 mg/Nm3
|
45 s
|
| NOrel / NOxrel - mass concentration of nitrogen oxides relative to O2 |
calculated from volumetric concentration of NO and O2
|
0...
|
1 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
1 mg/Nm3
|
45 s
|
| NO2 - volumetric concentration of nitrogen dioxide |
electrochemical gas sensor
|
0..1000 ppm
|
1 ppm
|
±5 ppm or 5 % rel.
|
5 ppm
|
45 s
|
| NO2mg - mass concentration of nitrogen dioxide |
calculated from volumetric concentration of NO2
|
0...
|
2 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
2 mg/Nm3
|
45 s
|
| NO2rel - mass concentration of nitrogen dioxide relative to O2 |
calculated from volumetric concentration of NO2 and O2
|
0...
|
2 mg/Nm3
|
±10 mg/Nm3 or 5 % rel.
|
2 mg/Nm3
|
45 s
|
| SO2 - volumetric concentration of sulphur dioxide |
electrochemical gas sensor
|
0..5000 ppm
|
1 ppm
|
±5 ppm or
5 % rel.
|
5 ppm
|
45 s
|
| SO2mg - mass concentration of sulphur dioxide |
calculated from volumetric concentration of SO2
|
0...
|
3 mg/Nm3
|
±15 mg/Nm3 or 5 % rel.
|
3 mg/Nm3
|
45 s
|
| SO2rel - mass concentration of sulphur dioxide relative to O2 |
calculated from volumetric concentration of SO2 and O2
|
0...
|
3 mg/Nm3
|
±15 mg/Nm3 or 5 % rel.
|
3 mg/Nm3
|
45 s
|
|
Gases measured with optional IR sensors |
| N2O - nitrous oxide volumetric concentration |
NDIR sensor
|
0..2000 ppm
|
1 ppm
|
±5 ppm or 5 % rel.
|
5 ppm
|
45 s
|
| CO2 - carbon dioxide volumetric concentration |
NDIR sensor
|
0..25 %
0...100 %
|
0.01 %
|
+/-3 % rel., or 0.5 % abs. |
0.20 %
|
45 s
|
| CH4 - Methane, volumetric concentration |
NDIR sensor
|
0...5 %
0...100 %
|
0.01 %
0.1 %
|
+/-3 % rel., or 0.5 % abs. |
0.20 %
|
45 s
|
|
Other measured values |
| Tgas - flue gas temperature |
Thermocouple
|
-10..1000°C
|
1°C
|
± 2 °C or 1.5 % rel.
|
1 °C
|
30 s
|
| Tamb - ambient temperature |
Thermistor
|
-10..100°C
|
1°C
|
± 1 °C
|
1 °C
|
30 s
|
| Pressure |
DMS bridge
|
-20 hPa ... +20 hPa
|
0.1Pa
|
±2 Pa or 5 % rel.
|
1 Pa
|
10 s
|
| Diff. Pressure |
DMS bridge
|
-20 hPa ... +20 hPa
|
0.1Pa
|
±2 Pa or 5 % rel.
|
1 Pa
|
10 s
|
| Soot/smoke test |
Bacharach method
|
0..9
|
0.5
|
0.5
|
0.5
|
|
|
|
| TI (CO/CO2-Toxic Index) |
calculated
|
0...0.01
|
0.0001
|
5 % rel.
|
0
|
10 s
|
| Lambda - excess air number |
calculated
|
1..10
|
0.01
|
2 % rel.
|
0
|
10 s
|
| qA - combustion losses |
calculated
|
0..100 %
|
0.1 %
|
2 % rel.
|
0 %
|
10 s
|
| Eta - efficiency |
calculated
|
0..120 %
|
0.1 %
|
2 % rel.
|
0 %
|
10 s
|